
TABLE 18-1
Specifications of the ACR CT Accreditation Phantom
| Phantom Construction | |
| Matrix material | Solid Water, 0 ± 5 HU |
| Length | 16 cm (6.30 inches) |
| Diameter | 20 cm (7.88 inches) |
| Weight | 5.3 kg (11.75 pounds) |
| Imbedded Test Objects | |
| Water equivalent linearity rod | Solid Water, 0 HU |
| Bone equivalent linearity rod | 955 HU bone tissue-equivalent material |
| Acrylic linearity rod | Cast acrylic |
| Polyethylene linearity rod | Low-density polyethylene |
| Low-contrast module matrix | Ciba Geigy CB4 epoxy or equivalent |
| Low-contrast rods | Ciba Geigy CB4 epoxy (density adjusted to yield 6 ± 0.5 HU difference) or equivalent |
| Tungsten carbide beads | 0.011 in diameter grade 25 tungsten carbide beads |
| Line pair material | 6061 aluminum and polystyrene |
| Steel beads | 1.00 mm grade 25 chrome steel balls |
| Intramodule homogeneity | The mean ROI values within any module, test objects excluded, can differ by no more than 2 HU |
| Intraphantom homogeneity, modules 1, 3, and 4 | The average CT number of a module must meet the requirements of 0 ± 5 HU |

(Courtesy Gammex Inc., Middleton, Wis.)


TABLE 18-2
Acceptance Criteria or Tolerance Limits for Three CT QC Tests from one International Organization (IAEA) and Two National Organizations (ACR and RPB-HC)
| QC Tests | ACCEPTANCE CRITERIA OR TOLERANCE LIMITS | ||
| ACR | IAEA | RPB-HC | |
| CT numberaccuracy | 0 ± 7 HU (±5 HU preferred) | 0 ± 5 HU | 0 ± 4 HU |
| CT number uniformity | The center CT number must be between ±7 HU (±5 HU preferred) Adult abdomen protocol must be used | ±10 HU | CT number for water must not be greater than ±5 HU from the center of the phantom to the periphery |
| Radiation dose | The CTDIw should not exceed 35 mGy for the adult abdomen protocol and CTDI | ±20% of manufacturer’s specification | CTDIw, which must be within ±20% of the established baseline values and the manufacturer’s specifications; it is highly recommended to strive for an agreement with manufacturers’ specification of ±10% |

TABLE 18-3
Acceptance Criteria for the ACR CT Accreditation QC Tests Using the ACR Phantom

Data from ACR. (2012). 2012 computed tomography manual: radiologic technologists’ section. <http://www.acr.org/Quality-Safety/accreditation/CT> Accessed February 2015.

















