Page 90 
  Page 91 

Chapter 12 Ultrasound safety

CONTENTS

Introduction 91
Why are we interested? 92
Operating modes and their potential risk 93
Who regulates ultrasound acoustic output limits? 94
Output display standards 94
Biological effects of ultrasound 94
Prudent use – the ALARA principle 97
General guidelines 97
Machine controls which influence hazard 98
Epidemiological evidence 98

LEARNING OBJECTIVES

1. Outline the parameters used to describe the strength of an ultrasound beam.
2. Discuss the significance of different operating modes and their potential risk of producing biological effects.
3. Be aware of current diagnostic ultrasound output limits.
4. Describe the thermal and non-thermal physical mechanisms that can cause biological effects.
5. Explain the thermal index (TI) and mechanical index (MI) and understand the significance of these in relation to causing a potential biological hazard.
6. Be aware of the general guidelines offered by ultrasound organizations and societies.
7. Discuss the influence of the machine controls.
8. Outline the ways to limit exposure by adopting the ALARA principle.
9. Have a basic appreciation of epidemiological evidence.

INTRODUCTION

There has been a considerable and rapid growth in the use of ultrasound as a diagnostic imaging tool over the past 50 years since it was first used to image the unborn fetus. This rapid growth has resulted in millions of people worldwide being scanned every year. Changes and advances in technology are leading to increasing power levels being used to obtain diagnostic information. The interaction of ultrasound with biological tissue can result in effects which may cause heating, interfere with the normal functioning of cells, and cause structural damage. The potential damaging nature of ultrasound leads to concern over its safety. All operators therefore need to have an understanding of the safety implications in order to practice safely.

  Page 92 

WHY ARE WE INTERESTED?

Ultrasound is a mechanical form of energy which interacts with the biological tissue through which it travels. Millions of women have their pregnancy routinely scanned with ultrasound and any adverse effects are more likely to damage the rapidly developing cells found in embryo, fetus, and neonate.

New research is continually being carried out, but variable parameters of ultrasound exposure and mechanisms of interaction with tissue lead to problems in determining when safe levels are being breached.

Important Parameters

Amplitude, power, and intensity are parameters used to describe the strength of an ultrasound beam.

Amplitude

Amplitude is a measure of a wave’s magnitude of oscillation, that is, the magnitude of the maximum disturbance in the medium, and in ultrasound often refers to the maximum variation (see Fig. 12.1). Amplitude is measured in units of pressure: MPa (megapascals).

image

Fig. 12.1 Demonstrating the varying pressure amplitude of an oscillating acoustic wave

Power (P)

Power in ultrasound describes the rate at which energy is generated and transferred by the acoustic wave per unit of time. Power is measured in watts (W) and milliwatts (mW), one milliwatt being one thousandth of a watt. Power is proportional to the square of the amplitude, i.e. if the power is doubled then the amplitude is quadrupled.

Intensity (I)

Intensity is the rate at which energy passes through the unit area and is an important quantity when discussing bioeffects and safety.

The average intensity is equal to the power of an ultrasound beam, normally expressed in mW, divided by the cross-sectional area of the beam, expressed in cm2. Units of intensity for diagnostic ultrasound are typically expressed in mWcm−2.

From the equation in Figure 12.2 we can identify that the intensity of an ultrasound beam is directly proportional to its power, i.e. if beam power increases, then intensity increases and, conversely, if beam power decreases, the intensity decreases.

image

Fig. 12.2 Diagram showing a typical ultrasound beam profile and a cross-sectional area. Cross-sectional area varies with depth and is smallest at the focus where the intensity is the highest

In addition, we can see that intensity is also inversely proportional to the beam area, i.e. if the beam area decreases, then the beam intensity increases and, conversely, if the beam area increases, the beam intensity decreases.

The maximum intensity along an ultrasound beam lies at the focus (the narrowest part of the beam)where all the power is concentrated into a small cross-sectional area.

  Page 93 

The intensity within an ultrasound beam also varies from point to point across the beam (spatial considerations), as demonstrated in Figure 12.3. Two values of intensity can de defined:

Spatial peak intensity (ISP) – is the highest intensity and is found at the center of the beam
Spatial average intensity (ISA) – is the average intensity across the beam and takes account of the variation across the width of the beam.
image

Fig. 12.3 Demonstrating the variation of intensity across an ultrasound beam. The spatial intensity across the beam is at its highest at the center, tailing off towards the edges

Because of the pulsed nature of ultrasound, the intensity of the ultrasound beam also varies over time (temporal considerations). Figure 12.4 shows the relevant times for three intensities:

Temporal peak intensity (ITP) – is the highest intensity found in a pulse and is not averaged over time
Pulse average intensity (IPA) – is the intensity averaged over the pulse duration. Lies in between ITP and ITA
Temporal average intensity (ITA) – takes account of the ‘dead’ time between pulses and is averaged over the pulse repetition period (PRP). It is the lowest value.
image

Fig. 12.4 Demonstrating the three time-dependent intensities

These spatial and temporal variations within the ultrasound beam result in the fact that there are a number of ways of defining intensity, as detailed below:

Highest intensity image ISPTP – Spatial peak-temporal peak (SPTP)
ISATP – Spatial average-temporal peak (SATP)
ISPPA – Spatial peak-pulse average (SPPA)
ISAPA – Spatial average-pulse average (SAPA)
ISPTA – Spatial peak-temporal average (SPTA)
Lowest intensity ISATA – Spatial average-temporal average (SATA)

ISPTA is the measure most associated with temperature rises.

OPERATING MODES AND THEIR POTENTIAL RISK

For any operating mode there is a large variation of output powers and intensities. The output data for all modern day ultrasound machines can be sourced from the operator’s manual.

Exposures used in Doppler modes such as spectral pulsed Doppler and color flow imaging are higher than for B and M modes.

B modes

Uses the lowest output power and intensities and is generally considered safe in all applications.

Pulsed Doppler

The longer pulses, higher powers, and pulse repetition rates typically used in pulsed Doppler result in higher average intensities compared to B-mode imaging and therefore there is an increased potential of producing a biological effect, particularly from ultrasound-induced heating.

  Page 94 

Dwell time, i.e. the length of time that the ultrasound beam is fixed on a specific tissue area, is an important factor when considering the potential for heating. This is particularly relevant for spectral pulsed Doppler where the beam is held in a fixed position during an investigation; this leads to a further increase in temporal average intensity and therefore increased risk of causing a temperature rise.

Color Flow Mapping and Power Doppler Imaging Modes

These modes involve some beam scanning, and so generally have a heating potential that is somewhere between that of B-mode and that of spectral pulsed Doppler.

Table 12.1 summarizes data collected from a survey conducted by Henderson in 1997. Here you can see that the highest intensities are produced when using pulsed Doppler modes.

Table 12.1 Values of the UK survey of equipment in clinical use (from Henderson et al 1997).

image

WHO REGULATES ULTRASOUND ACOUSTIC OUTPUT LIMITS?

Surprisingly enough, these are not regulated by a medical body. Regulations governing the output of diagnostic ultrasound have been largely set by the USA’s Food and Drug Administration (FDA). In 1991 the FDA relaxed its regulations and increased the permissible acoustic output and intensity levels used to scan early pregnancy to almost eight times over the level that had been previously allowed (from 94 to 720 mWcm−2). Table 12.2 shows these changes in acoustic output levels in relation to the range of diagnostic areas.

Table 12.2 Shows the changes to maximum allowable output exposure following the changes stipulated by the US FDA. ODS; output display standards.

APPLICATION BEFORE 1991 (PRE ODS) INTENSITY LIMIT (ISPTA) (mWcm−2) AFTER 1991 WITH ODS INTENSITY LIMIT (ISPTA) (mWcm−2)
Fetal, neonatal imaging 94 720
Cardiac 430 720
Peripheral vascular 720 720
Ophthalmology 17 50

However, for the manufacturers to utilize these increased output levels and higher intensities, it was agreed that their equipment should provide an on-screen display relating to the acoustic output. Its purpose is to inform the user of the potential risk for an examination to produce a biological effect. This shifted the onus of safety to the end user who is now responsible for making a risk/benefit assessment. This requires the ultrasound user to understand the potential for biological effects and practice to ensure exposure times are kept to a minimum.

OUTPUT DISPLAY STANDARDS

In 1992, the American Institute of Ultrasound in Medicine (AIUM) and the National Electrical Manufacturers Association (NEMA) defined a set of output display standards (ODS). For each ultrasound examination, a real-time on-screen display indicates the risk of producing biological effects. ODS is based on two indices which give two main measures of output:

TI = Thermal effects (related to average intensity)
MI = Cavitation – non-thermal (related to peak pressure).

This enables users to be aware of output levels during scanning.

These increases in output levels have also driven some of the improvements and developments in diagnostic ultrasound that we mentioned earlier, to the extent that modern ultrasound equipment is capable of producing measurable biological effects in tissues (Barnett 1998, Barnett & Kossoff 1992).

BIOLOGICAL EFFECTS OF ULTRASOUND

Physical Mechanisms

Clinically we know ultrasound energy can be utilized in lithotripsy by using a focused, high-intensity acoustic pulse to destroy a kidney stone. These devices use energies much higher than those utilized in diagnostic ultrasound, however it can be easily demonstrated that there is potential for destruction.

  Page 95 

There are two distinct mechanisms that can cause biological effects on tissue when exposed to ultrasound. These are:

Thermal (heating)
Non-thermal effects (cavitation).

Thermal effects

As the ultrasound beam travels through tissue, some of its energy is lost through absorption. This absorbed energy is converted into heat which in turn produces a temperature rise. In most soft tissues, the rate at which energy is absorbed depends on the:

attenuation coefficient of the tissue
operating frequency
power/intensity of the ultrasound beam
length of operating time (exposure time).

The thermal effect is highest in tissues with a high absorption coefficient such as bone, and is low where there is little absorption such as amniotic fluid (water).

The potential of any temperature rise also depends on the thermal characteristics of the tissue to dissipate this heat to surrounding areas, the ultrasound intensity, and the length of time for which the tissue volume is scanned.

Frequency is another factor and the higher the operating frequency, the higher the absorption and potential to cause heat. This is particularly significant when using endocavity probes which use higher frequencies to improve image quality.

The ultrasound intensity varies along the ultrasound beam and is greatest at the focus where the beam is at its narrowest. The intensity can be changed by other operator controls such as power output, scan depth, and mode of operation, i.e. using color and Doppler modes.

Heat can also be produced at the transducer surface which can be directly transferred to the tissue it is in contact with. This can be particularly significant when using transvaginal probes to investigate a pregnancy during the first eight weeks postconception.

The World Federation for Ultrasound in Medicine and Biology (WFUMB; Barnett 1998), following a review of the literature on the effects of temperature elevation on animal fetuses, concluded that a temperature rise of no more than 1.5 °C may be used clinically without reservation, however, exposures that elevate embryonic and fetal temperatures higher than 4 °C for 5 minutes or more should be considered as potentially hazardous.

Thermal index (TI)

The thermal index (TI) as defined by AIUM/NEMA is the ratio of the acoustic power produced by the transducer (W) to the power required to raise the temperature in tissue by 1 °C (Wdeg). This is mathematically expressed in Equation 1:


image


Equation 1: Formula for thermal index.

(Derated values compensate for attenuation by soft tissue. Acoustic output measurements are usually performed in a water tank. To take into account the attenuation by soft tissue in practice, these values are adjusted, i.e. ‘derated’, using a value of attenuation of 0.3 dB/cm/MHz.)

A thermal index of 1 indicates the acoustic power achieving a temperature increase of 1 °C. A thermal index of 2 has doubled power but would not necessarily indicate a peak temperature rise of 2 °C. The algorithms in the machine which calculate TI are based on many assumptions and take into account a worst case scenario such as low levels of beam attenuation, a stationary beam, and fairly long exposure times.

There are several classifications of thermal indices:

TIS – this is the thermal index applied when you are scanning through a soft-tissue structure and will probably be the one you most often encounter
TIB – this is the thermal index which applies if bone is at or near the focus of the transducer. Remember, because of its high acoustic absorption, there is an increased potential for heating which will be reflected in the values of TI displayed. TIB is most appropriate for fetal ultrasound during the second or third trimester pregnancy scan where the highest temperature increase would be expected occurring at the soft tissue/bone interface.
TIC – this is the thermal index applied when scanning cranial bone. Here, the bone is at or very near the skin surface.

Cavitation effects

Ultrasound produces an oscillating pressure wave which propagates through tissues. This propagating pressure wave can cause micro-bubbles within tissues to form, grow, oscillate in size and, at sufficiently high intensities and pressure, cause these bubbles to collapse. Small laboratory mammalian experiments have demonstrated this effect in lung and intestine, where small gas bubbles can be present. The most important sources of cavitation are those introduced during contrast studies.

Above diagnostic output levels, cavitation plays an important role in the destruction of kidney stones in shock wave lithotripsy.

There are two forms of cavitation:

Stable (also known as non-inertial)
Unstable (also known as transient or inertial).

Stable cavitation (non-inertial)

This form of cavitation is the process where small micro-bubbles in a medium are forced to oscillate in the presence of the ultrasound. The cavitation bubbles go through phases of expansion and contraction as they oscillate with the varying ultrasound pressure wave. Generally, this form of stable avitation is regarded as being safe. Figure 12.5 graphically demonstrates the phases of bubble compression and contraction through the varying ultrasound pressure wave.

image

Fig. 12.5 Demonstrating the phases of bubble compression and expansion through the varying ultrasound pressure wave. Size of bubble is smallest when compressed during the peak positive pressure and is largest during the peak negative or rarefaction pressure

  Page 97 

Unstable cavitation (transient or inertial)

This type of cavitation has the greatest potential to damage tissue. Here, the intensity of the ultrasound field is high enough to cause the oscillations of the micro-bubbles to become so great that the bubbles collapse, generating high pressures and temperatures to the localized area, increasing the potential to cause biological damage. This effect has been demonstrated in small mammalian animal experiments causing lung capillary bleeding at pressure thresholds of 1 MPa which is well within the diagnostic range. Although this process has been reported in mammalian experiments, it has not been observed in humans. However, knowledge that there is a potential risk and hazard for biological damage due to this form of destructive cavitation is especially pertinent in early pregnancy where the tissues are more susceptible to this form of damage.

The use of contrast agents in the form of stabilized gas bubbles increases the probability of cavitation.

Mechanical index (MI)

The mechanical index (MI) is related to the likelihood of cavitation being produced and is defined as the peak rarefactional pressure (Pr) (negative pressure) divided by the square root of the ultrasound frequency:


image


Equation 2: Formula for calculating mechanical index.

The value of Pr (negative pressure) is significant here and from Equation 2 we can see that the value of MI is directly proportional to Pr. This means if Pr doubles then MI doubles and, conversely, if Pr is halved then MI is halved. Also we can see that frequency is inversely related to MI which means as frequency increases the value of MI actually decreases.

The FDA ultrasound regulations allow a mechanical index of up to 1.9 to be used for all applications except ophthalmic (maximum 0.23). The used range varies from 0.05 to 1.9.

It is stipulated that the values of the TI and MI must be displayed if the ultrasound system is capable of exceeding an index of 1.

The TI and MI are not absolute values but represent a rough approximation of the risk of creating certain biological effects. The higher the index value, the higher the potential of a biological effect.

PRUDENT USE – THE ALARA PRINCIPLE

Changes in power output and increased use of Doppler ultrasound, together with a change in regulations governing output, mean that the onus of safety is placed on the operator. As a consequence, every measure should be taken to ensure safe practice.

All ultrasound users should regularly check both thermal and mechanical indices while scanning and should adjust the machine controls to keep them as low as reasonably possible without compromising the diagnostic value of the examination. Where low values cannot be achieved, examination times should be kept as short as possible.

The ALARA (‘As Low As Reasonably Achievable’) principle should be applied and maintained. Minimize any possible risks by minimizing exposure times and exposure intensity.

GENERAL GUIDELINES

There are many safety guidelines issued by several national and international ultrasound societies and organizations to assist all those who use diagnostic ultrasound so that they are able to make informed decisions about ultrasound safety. These guidelines are based on the best scientific information available at the time.

A list of these organizations is presented in Table 12.3 and further information can be accessed through their Web sites.

Table 12.3 List of ultrasound organizations and societies which offer safety guidelines

ORGANIZATION/SOCIETY WEB SITE
World Federation For Ultrasound in Medicine and Biology (WFUMB) http://www.wfumb.org
European Federation of Societies for Ultrasound in Medicine and Biology (EFSUMB) http://www.efsumb.org
British Medical Ultrasound Society (BMUS) http://www.bmus.org
American Institute of Ultrasound in Medicine (AIUM) http://www.aium.org
Australasian Society for Ultrasound in Medicine http://www.asum.com.au
  Page 98 

MACHINE CONTROLS WHICH INFLUENCE HAZARD

The exposure intensity and therefore any possible associated temperature rise are affected by many operator-dependent controls. The operator should understand the influence of the machine controls

Initial Power Setting/Output

Ultrasound machines are normally set up for all application-specific settings. This sets a default acoustic output which should be adequate to perform each examination. Application presets for obstetrics should default the machine’s output to a low setting. The operator should only increase the output power setting if it is necessary and required to improve image quality.

Mode of Operation

B mode generally has the lowest power output and intensity. M mode, color flow imaging, and spectral Doppler have higher outputs which can cause more heating at the site of examination. Examinations should use B mode in the first instance and employ the use of color and spectral Doppler only when and if required.

Exposure Time

The overall examination should be kept as short as possible to obtain the required level of diagnostic information.

Stationary Probe

The probe should not be held in any fixed position for any longer than necessary. Operators should keep the transducer moving so that specific tissues are not subjected to long periods of stationary beam exposure. Use freeze frame and cineloop which allow images to be reviewed without continuing exposure.

Other Machine Controls

The intensity (and temperature rise) is highly dependent on other scanner settings such as:

transmitted frequency used
depth of examination
focus.

The operator should be aware of and monitor changes to the values of TI and MI in response to changes in machine control settings and transducer combinations.

Table 12.4 summarizes the recommended guidelines issued by the British Medical Ultrasound Society (BMUS) in 2000 for thermal and mechanical indices. Full details of this can be accessed on their Web site.

Table 12.4 Recommended guidelines for values of MI and TI issued by BMUS

  BIOLOGICAL EFFECT ACTION BY OPERATOR
MI > 0.3 Possibility of minor damage to neonatal lung or intestines Reduce exposure times as much as possible
MI > 0.7
Theoretical risk of cavitation
Risk of cavitation when conducting contrast agent studies
Reduce exposure times as much as possible
TI > 0.7 Increased risk of heating Overall exposure time of an embryo or fetus should be restricted
TI > 1 Increased risk of heating
Eye scanning not recommended
Overall exposure time of an embryo or fetus should be restricted
TI > 3 Significant risk of heating Scanning of an embryo or fetus not recommended

EPIDEMIOLOGICAL EVIDENCE

Epidemiology is the study of how often diseases occur in different groups of people and why. A key feature of epidemiology is the measurement of disease outcomes in relation to a population at risk.

  Page 99 

As a consequence of the rapid development of ultrasound as a diagnostic tool, the general population has been and is being exposed to more ultrasound energy than ever before. This is especially true for exposures of the fetus.

In the case of ultrasound, epidemiological studies have been used to search for risks associated with previous exposure to diagnostic ultrasound of fetuses (in utero) in early pregnancy as it is known that the developing human embryo or fetus is particularly sensitive to the ultrasound beam.

The outcomes of published epidemiology research evidence have suggested associations between ultrasound exposure and factors such as reduced birth weight, delayed speech development, dyslexia, and non-right handedness. However, some of this evidence has been conflicting and the reliability of these outcomes has since been questioned.

To date, diagnostic ultrasound has not been proved to be unsafe in humans. It is important to note that much of this research work was conducted before output levels were relaxed by the FDA in 1991 and that there is limited data for modern powerful diagnostic equipment using Doppler, harmonic imaging, and contrast agents. Ongoing research into bioeffects is necessary to reaffirm that ultrasound is still a safe modality.

Continuous vigilance is necessary, particularly in areas of concern such as the use of pulsed Doppler in the first trimester.

SUMMARY

To date, diagnostic ultrasound has not been proved to be unsafe in humans
Epidemiology data have several limitations:
Limited data for modern powerful diagnostic equipment
Ongoing research into bioeffects still necessary
Using pulsed Doppler, harmonic imaging, contrast agents
Output regulations relaxed in 1991 for obstetric scans (> 7 times increase)
Operator responsible for risk/benefit analysis
Many machine controls have a significant effect
Displayed MI and TI remove some guesswork for end users
Knowledge, understanding, and training important
Prudent use is practiced – ALARA principle
Many guidelines offered to end users

References

. Conclusions and recommendations on thermal and non thermal mechanisms for biological effects of ultrasound. Barnett SB, editor. WFUMB (World Federation for Ultrasound in Medicine and Biology) Symposium on Safety for Ultrasound in Medicine, 24. Ultrasound in Medicine and Biology. 1998:1-55.

Issues and recommendations regarding thermal mechanisms for biological effects of ultrasound. In: Barnett SB, Kossoff G, editors. WFUMB (World Federation for Ultrasound in Medicine and Biology) Symposium on Safety and Standardisation in Medical Ultrasound. Ultrasound in Medicine and Biology; 1992:18. special issue

British Medical Ultrasound Society (BMUS). Guidelines for the safe use of diagnostic ultrasound equipment. www.bmus.org/ultras-safety/us-safety03.asp. Online. Available:

Henderson J, Whittingham TA, Dunn T. A review of the acoustical output of modern diagnostic ultrasound equipment. BMUS Bulletin. 1997;5:10-14.