Page 50 
  Page 51 

Chapter 9 Artifacts

CONTENTS

Assumptions made by ultrasound equipment 51
Reverberation 52
Acoustic shadowing 53
Acoustic enhancement 54
Edge shadowing 55
Beam width artifact 56
Slice thickness artifact 56
Side lobe artifact 58
Mirror image artifact 59
Double image artifact 59
Equipment-generated artifacts 60

LEARNING OBJECTIVES

1. List the basic assumptions made by ultrasound equipment which gives rise to artifacts.
2. Describe the appearances of the more common artifacts encountered during ultrasound scanning.
3. Explain the causes of the commonly arising artifacts.

An ultrasound artifact is a structure in an image which does not directly correlate with actual tissue being scanned. Artifacts assume different forms, including:

structures in the image that are not actually present
objects that should be represented but are missing from the image
structures which are misregistered on the image.

Where an operator is not aware of the presence of artifacts, it is possible that pathology will not be demonstrated and will therefore not be diagnosed (false negative result), or that pathology will be diagnosed where it does not exist (false positive result).

Understanding the basic mechanisms behind ultrasound artifacts, recognizing situations in which they are likely to arise, and becoming familiar with their appearance all help to eliminate misdiagnoses that may otherwise occur.

ASSUMPTIONS MADE BY ULTRASOUND EQUIPMENT

Ultrasound machines have to make certain assumptions in order to operate. These are:

  Page 52 
The speed of sound propagation is the same in all tissue types (usually equipment assumes this to be equal to 1540 m/s)
The ultrasound beam travels in straight lines
The time taken after emission of a pulse for an echo to return to the transducer from an interface is directly related to the distance of the interface from the transducer
Attenuation of sound in tissue is uniform
All echoes detected by the transducer have arisen from the central axis of the beam
An echo generated at an interface returns directly in a straight line to the transducer without generating any secondary echoes
The intensity or strength of the returning echo is proportional to the density of the reflector generating the echo.

These assumptions are often incorrect and will give rise to appearances which do not correspond to actual anatomy or tissue in the scan plane. It is these appearances that we term artifacts.

Several possible artifactual appearances may occur:

There may be areas incorrectly showing absence of echoes
Echoes may appear in areas which do not represent actual interfaces
Separate structures may appear joined
A single structure may appear as separate structures
The brightness or textural appearance of echoes may not correlate with actual tissue structure
Thin structures may appear thicker than reality or vice versa.

There are a multitude of different artifacts which may occur, but the more commonly occurring artifacts which will be discussed in this chapter are:

reverberation
acoustic shadowing
acoustic enhancement
edge shadowing
beam width artifact
slice thickness artifact
side lobe artifacts
mirror image
double image
equipment-generated artifacts.

REVERBERATION

This is the production of spurious (false) echoes due to repeated reflections between two interfaces with a high acoustic impedance mismatch. The echo from the interface is received by the transducer and displayed on the image. Some of the energy in the returned echo is reflected at the transducer face, and returns to the reflecting interface as if it was a weak transmitted pulse, returning as a second echo (reverberation). As the time taken for the second echo to arrive is twice that taken by the first echo, the image will display it at twice the depth. This sequence of reflection and transmission can occur many times, with the third echo taking three times as long to return to the transducer and being displayed at three times the depth, and so on. The reverberation echoes will be equally spaced because the time for each additional echo is a multiple of the time of return of the first echo. These reverberation echoes will be strong because of the high acoustic mismatch (see Fig. 9.1).

image

Fig. 9.1 Repeated reflections between two interfaces cause reverberation of the displayed echoes

This artifact will often be seen:

at the skin–transducer interface when the sound reverberates between the subcutaneous fat/muscle layer and the transducer (see Figs 9.2 and 9.3)
behind bowel gas when the sound reverberates between the gas surface and the transducer (also known as ringdown) (see Fig. 9.4).
image

Fig. 9.2 Reverberation occurring at the skin–transducer interface

image

Fig. 9.3 Reverberation arising at the skin–transducer interface and appearing in the bladder

image

Fig. 9.4 Reverberation occurring between bowel gas and transducer, appearing in the image distal to the bowel gas, and known as ringdown

The artifact can be differentiated from real echoes due to the lack of breathing movement occurring.

  Page 53 

To help to eliminate this artifact it may be useful to:

increase the amount of gel used
use a stand-off pad (such as a commercially produced gel pad or a bag of saline)
reduce the gain
move the position of the transducer.

ACOUSTIC SHADOWING

This appears as an area of low amplitude echoes (hypoechoic or anechoic) behind an area of strongly attenuating tissue. It is caused by severe attenuation of the beam at an interface, resulting in very little sound being transmitted beyond (see Fig. 9.5). The attenuation can be due to either absorption or reflection of the sound waves, or a combination of the two.

image

Fig. 9.5 Acoustic shadowing posterior to a strongly attenuating gallstone

Acoustic shadowing will occur at interfaces with a large acoustic mismatch such as:

soft tissue and gas (where the degree of attenuation will be due to reflection of 99.9% of the beam)
soft tissue and bone or calculus (where the attenuation is due to a combination of approximately 40% reflection of the beam and 60% absorption by the bone).

It is often possible to differentiate diagnostically between structures such as gas in the duodenum or calculi in a small contracted gallbladder, by looking at the type of acoustic shadowing. Often shadowing in this area can be ambiguous and lead to an inaccurate diagnosis, but careful examination will reveal different types of shadowing. If the shadowing contains reverberation (ringdown) echoes, it is likely to be gas (see Fig. 9.6a & b). If the shadowing is clear shadowing, absent of any ringdown, it is likely to be a calculus (see Fig. 9.7 a & b).

image image

Fig. 9.6a & b Acoustic shadowing occurring in the region of the gallbladder (suggesting possible biliary calculi), but as it contains ringdown, it is more likely to be caused by gas in the adjacent bowel

image image

Fig. 9.7a & b Acoustic shadowing posterior to biliary calculi due to a combination of reflection and absorption

  Page 54 

ACOUSTIC ENHANCEMENT

This artifact appears as a localized area of increased echo amplitude behind an area of low attenuation. On a scan it will appear as an area of increased brightness, and can commonly be seen distal to fluid-filled structures such as the urinary bladder, the gallbladder, or a cyst.

  Page 55 

The artifact arises due to the application of the time-gain compensation (TGC) to areas of low attenuating structures such as fluid. It is caused by the low level of attenuation of the beam as it passes through fluid relative to the greater attenuation of the beam in the adjacent more solid tissue (see Fig. 9.8). As the echoes pass beyond the area, they will be of higher amplitude than the surrounding tissue because they have been amplified unnecessarily. The high amplitude brightness therefore does not relate to any inherent scattering or reflectivity properties of the tissue, but arises because the equipment assumes a uniform rate of attenuation across the entire image (see Fig. 9.9).

image

Fig. 9.8 Acoustic enhancement posterior to a low attenuating cyst

image

Fig. 9.9 Acoustic enhancement posterior to a cyst

This artifact can often be a useful diagnostic aid, particularly when scanning a soft-tissue mass or cyst containing low-level echoes. These echoes may often cause the structure to disappear in the image as it blends into the surrounding echo pattern. The observant operator will often notice the echo enhancement, resulting in a closer examination of the area above this, and the mass or cyst will then be delineated (see Fig. 9.10).

image

Fig. 9.10 Acoustic enhancement posterior to a soft tissue mass

EDGE SHADOWING

A combination of refraction and reflection occurring at the edges of rounded structures (and when the speed of sound is different from that in surrounding tissue) will result in an edge shadowing artifact.

Edge shadowing arises due to refraction of the beam caused by both the curvature of the rounded edges and the difference in speed of the two materials. When the ultrasound beam reaches the rounded edge of a structure, reflection will occur, with the angle of incidence equal to the angle of reflection. The outer part of the beam will be totally reflected, but the remainder of the beam passes through the rounded structure and is refracted (deviated from its original path) (see Fig. 9.11). This combination of reflection and refraction of the beam at the edge of a rounded structure results in a thin strip of tissue behind the edge not being insonated and causes a shadow. These shadows are narrow and occur directly distal to the margins of the rounded structure, such as either a cyst (see Fig. 9.12) or a soft-tissue mass (see Fig. 9.13). The shadowing should normally be readily identified as an artifact, however it is possible that it may be falsely diagnosed as areas of calcification within an organ.

image

Fig. 9.11 Edge shadowing

image

Fig. 9.12 Edge shadowing posterior to a cyst

image

Fig. 9.13 Edge shadowing posterior to a soft-tissue mass

  Page 56 

BEAM WIDTH ARTIFACT

All the echoes returning to the transducer will have arisen from across the full width of the beam, which can vary by several millimeters. As the beam sweeps through the patient, a point reflector will generate an echo for as long as it remains in the beam, and the reflector will be represented as a line in the display rather than a dot. The length of the line will therefore represent the beam width at that depth. This artifact can be demonstrated by scanning a point reflector in a phantom, where the display will clearly portray this as a line (Fig. 9.14).

image

Fig. 9.14 Scan of a phantom demonstrating that the point reflectors can be seen as points in the center of the image where the focal zone is set (and the beam is therefore narrowest), but the same points appear as lines where they lie outside the focal zone (and the beam is wider)

During routine scanning, the artifact can be seen when spurious echoes are displayed in an echo-free area, for example when reflections from bowel are generated by the edge of the beam and displayed inside the sagittal view of the urinary bladder, where the center of the beam is (see Figs 9.15 & 9.16).

image

Fig. 9.15 Beam width artifact

image

Fig. 9.16 Beam width artifact in a sagittal view of the bladder

Correct positioning of the focal zone will help to reduce this artifact. The focal zone is controlled by electronically narrowing the beam (see Chapter 5 on The ultrasound beam).

SLICE THICKNESS ARTIFACT

These occur due to the thickness of the beam, and are similar to beam width artifacts (see Chapter 5 on The ultrasound beam). However, they occur at 90° to the scan plane (see Fig. 9.17), with echoes from interfaces in front of and behind the assumed plane of origin appearing in the display.

image

Fig. 9.17 Beam width and slice thickness

  Page 57 

It is assumed that the echoes detected by the transducer are from a very thin slice of tissue. However, the reality is that this slice is actually composed of several slices of information and therefore echoes from interfaces on either side of the intended thin slice will be included in the displayed image (see Fig. 9.18). Increasing the slice thickness will increase the number of artifactual echoes in the display.

image

Fig. 9.18 Each image is composed of several slices of information added together

These artifacts will typically be seen in transverse views of the urinary bladder when structures adjacent to the slice through the bladder being scanned will be incorporated into the image. These echoes are then displayed as if they were arising from within the bladder (see Fig. 9.19). Although the appearance of this artifact is similar to the beam width artifact, the differentiating factor is that the reflector causing the slice thickness artifact will not be seen on the display.

image

Fig. 9.19 Slice thickness artifact in a transverse view of the bladder

This artifact is a result of inherent characteristics of the transducer, and apart from trying a different transducer, cannot be eliminated. New technology, however, is continuously resulting in narrower slice thicknesses.

  Page 58 

SIDE LOBE ARTIFACT

The energy within the ultrasound beam exists as several side lobes radiating at a number of angles from a central lobe (see Fig. 9.20). Echoes are generated by these side lobes in addition to the main lobe, but all the returning echoes are assumed by the transducer to have arisen from the central axis of the main lobe (see Chapter 5 on The ultrasound beam). Side lobe echoes will therefore be misregistered in the display. Because ultrasound beams are three-dimensional, side lobes exist not only in the scan plane but also in the entire 360° around the beam. The echoes generated by the side lobes are usually much lower in amplitude than the main beam and will therefore often not be seen, unless the echoes are returning from a strong reflector such as gas.

image

Fig. 9.20 Side lobes

This artifact can often be seen in areas such as the urinary bladder where the side lobes detect echoes from adjacent bowel and the equipment registers these echoes as if they have arisen from within the bladder. They may also arise within a cyst, where adjacent structures are portrayed as if arising from within the cyst (see Fig. 9.21). These appearances can give rise to a false diagnosis of septations within a cyst.

image

Fig. 9.21 Artifact in a cyst caused by information obtained by the side lobes and leading to possible misdiagnosis of a septated cyst

  Page 59 

MIRROR IMAGE ARTIFACT

These artifacts result in a mirror image of a structure occurring in an ultrasound display. They arise due to specular reflection of the beam at a large smooth interface. An area close to a specular reflector will be imaged twice, once by the original ultrasound beam and once by the beam after it has reflected off the specular reflector. Echoes return along the same path from the reflecting interface, back to the transducer. Because the equipment assumes all echoes arise from a straight beam, the reflected echoes are displayed in a line as if originating from below the specular reflecting surface (see Fig. 9.22).

image

Fig. 9.22 Mirror image artifact

Mirror image artifacts are most commonly seen where there is a large acoustic mismatch, such as a fluid–air interface. Typically this artifact can occur during the scanning of a full urinary bladder, when air in the rectum behind the bladder acts as a specular reflector and a mirror image of the bladder is displayed posteriorly. It will then have the appearance of a large cyst behind the bladder (see Fig. 9.23). It can also be seen when scanning the liver, and the diaphragm acts as a specular reflector. In this case the liver parenchyma is displayed not only below the diaphragm but also above it (see Fig. 9.24). When trying to determine whether these appearances represent pathology or artifact, it is important to recognize whether the echoes have the same appearance as the organ from which they have arisen.

image

Fig. 9.23 Mirror image artifact of the bladder imitating a cyst in the pelvis

image

Fig. 9.24 Mirror image artifact of the liver appearing above the diaphragm

(From Bates 1999, with permission of Churchill Livingstone.)

DOUBLE IMAGE ARTIFACT

This artifact is caused by refraction of the beam and may occur in areas such as the rectus abdominis muscle on the anterior abdominal wall. In the transverse plane the edges of the muscle act as a lens and cause the ultrasound beam to be refracted (i.e. deviated from its original path) and this causes a single structure to be interrogated by two separate refracted beams (see Fig. 9.25). Two sets of echoes will therefore be returned and these will cause display of two structures in the image. This results in, for example, two images of the transverse aorta side by side in the abdomen (see Fig. 9.26). Alternatively, a single gestation sac can be mirrored and appear as two gestation sacs side by side, leading to the erroneous diagnosis of a twin pregnancy. In order to establish whether these echoes are genuine, it is necessary to move the transducer slightly to one side to avoid the junction of the rectus abdominis muscles.

image

Fig. 9.25 Double image artifact caused by refraction of the beam

image

Fig. 9.26 Double image artifact of the aorta

  Page 60 

EQUIPMENT-GENERATED ARTIFACTS

Incorrect use of the equipment controls can lead to artifacts appearing. Misuse of controls such as the gain (see Fig. 9.27) or TGC (see Figs 9.28 & 9.29) can result in echoes being recorded as too bright or too dark. Care must be taken when setting these controls, to ensure an even brightness throughout the image. If too much gain is applied then the electronic noise, inherent in all systems, will also be amplified. This has the appearance of a fine overlay of low-level echoes in the image and will cause deterioration of the quality of the image, reducing the ability of the operator to correctly interpret. If too little gain is applied, this can lead to loss of relevant information, and incorrect diagnosis may occur.

image

Fig. 9.27 Too much gain causing artifactual echoes

image

Fig. 9.28 Incorrect use of TGC causing artifactual echoes

image

Fig. 9.29 Incorrect use of TGC causing artifactual appearances

  Page 61 

If the dynamic range control is incorrectly set, this can lead to an image which has too much contrast, and result in the loss of subtle echo information (see Fig. 9.30).

image

Fig. 9.30 The image has too much contrast resulting in the loss of subtle information

Blurring of a moving image can occur if the frame rate is too low or if the persistence is too high. It is important to ensure that the frame rate is capable of recording a moving structure at the correct speed.

Use of multiple focal zones can give rise to a prominent banding effect within the image. These are usually identifiable as being of electronic origin due to the well-defined margin of the bands.

SUMMARY

An ultrasound artifact is a structure in an image which does not directly correlate with actual tissue or structures
Artifacts occur due to false assumptions made by the equipment
Failure to recognize artifacts can lead to misdiagnosis
Many different artifacts occur during routine ultrasound scanning but the more common ones include:
Reverberation
Acoustic shadowing
Acoustic enhancement
Edge shadowing
Beam width artifact
Slice thickness artifact
Side lobe artifact
Mirror image
Double image artifact
Equipment-generated artifacts

Reference

Bates JA. Abdominal ultrasound, 2nd edn. Edinburgh: Churchill Livingstone, 1999.