Page 207 

Chapter 18 The quality of radiographic images and quality assurance

INTRODUCTION

The factors that can affect the quality of radiographic images depends on:

How the image was taken (radiographic technique)

What image receptor was used (film or digital)

How the visual image was created

Chemical processing (film)
Computer processing (digital).

The effects of poor radiographic technique are the same whatever type of image receptor is used. These technique errors have already been covered in detail in relation to the three main projections used in dentistry, namely: periapicals (Ch. 10), bitewings (Ch. 11) and panoramic radiographs (Ch. 17).

The creation of the visual digital image was described in Chapter 7, together with how computer software can be used to alter and manipulate the image with regards to contrast, brightness (degree of blackening), magnification, inversion, enhancement and pseudocolourization.

Creation of the black/white/grey image on film using chemical processing was also described in Chapter 7. This film-captured image can however be affected by many other factors. This chapter therefore is designed for revision, bringing together and summarizing from earlier chapters all these various factors. It also includes a quick reference section as an aid to fault-finding of film-captured images. Various film faults are illustrated together with their possible causes. This is followed by a section on quality assurance (QA) and suggested quality control measures.

IMAGE QUALITY

As mentioned in Chapter 1, image quality and the amount of detail shown on a radiographic film depends on several factors including:

Contrast

Image geometry

Characteristics of the X-ray beam

Image sharpness and resolution.

Contrast

Radiographic contrast, i.e. the final visual difference between the various black, white and grey shadows depends on:

Subject contrast

Film contrast

Fog and scatter.

Subject contrast

This is the difference caused by different degrees of attenuation as the X-ray beam is transmitted through different parts of the patient’s tissues. It depends upon:

Differences in tissue thickness

Differences in tissue density

Differences in tissue atomic number (photoelectric absorption ∝ Z3 (see Ch. 2))

Quality (voltage (kV)) or penetrating power of the radiation beam (kV↑, contrast↓ — kV kills contrast).

Film contrast

This is an inherent property of the film itself (see Ch. 6). It determines how the film will respond to the different exposures it receives after the X-ray beam has passed through the patient. Film contrast depends upon four factors:

The characteristic curve of the film

Optical density or degree of blackening of the film

Type of film — direct or indirect action

Processing.

Fog and scatter

Stray radiation reaching the film either as a result of background fog, or owing to scatter from within the patient, produces unwanted film density (blackening), and thus reduces radiographic contrast.

Image geometry

As mentioned and illustrated in Chapter 1, the geometric accuracy of any image depends upon the position of the X-ray beam, object and image receptor (film or digital) satisfying certain basic geometrical requirements:

The object and the film should be in contact or as close together as possible

The object and the film should be parallel to one another

The X-ray tubehead should be positioned so that the beam meets the object and the film at right angles.

Characteristics of the X-ray beam

The ideal X-ray beam used for imaging should be:

Sufficiently penetrating to pass through the patient, to a varying degree, and react with the film emulsion to produce good contrast between the various black, white and grey shadows (see earlier)

Parallel, i.e. non-diverging, to prevent magnification of the image (see Ch. 5)

Produced from a point source to reduce blurring of the image margins and the penumbra effect (see Ch. 5).

Image sharpness and resolution

Sharpness is defined as the ability of the X-ray film to define an edge. The main causes of loss of edge definition include:

Geometric unsharpness including the penumbra effect (see above)

Motion unsharpness, caused by the patient moving during the exposure

Absorption unsharpness — caused by variation in object shape, e.g. cervical burn-out at the neck of a tooth (see Ch. 21)

Screen unsharpness, caused by the diffusion and spread of the light emitted from intensifying screens (see Ch. 6)

Poor resolution. Resolution, or resolving power of the film, is a measure of the film’s ability to differentiate between different structures and record separate images of small objects placed very close together, and is determined mainly by characteristics of the film including:

type — direct or indirect action
speed
silver halide emulsion crystal size.

Resolution is measured in line pairs per mm.

PRACTICAL FACTORS INFLUENCING IMAGE QUALITY

In practical terms, the various factors that can influence overall image quality can be divided into factors related to:

The X-ray equipment

The image receptor — film or film/screen combination

Processing

The patient

The operator and radiographic technique.

As a result of all these variables, film faults and alterations in image quality are inevitable. However, since the diagnostic yield from radiography is related directly to the quality of the image, regular checks and monitoring of these variables are essential to achieve and maintain good quality radiographs. It is these checks which form the basis of quality assurance (QA) programmes.

  Page 209 

Clinicians using film need to be able to recognize the cause of the various film faults so that appropriate corrective action can be taken.

TYPICAL FILM FAULTS

Examples of typical film faults are shown below and summarized later in Table 18.1.

Table 18.1 Summary of common film quality problems and their possible causes.

image image

Film too dark (Figs 18.1 and 18.2)

Possible causes

Overexposure owing to:

Faulty X-ray equipment, e.g. timer
Incorrect exposure time selected

Overdevelopment owing to:

Excessive time in the developer solution
Developer solution too hot
Developer solution too concentrated

Fogging owing to:

Poor storage conditions:
* Allowing exposure to stray radiation
* Too warm
Old film stock used after expiry date
Faulty cassettes allowing ingress of light
Faulty darkroom/processing unit:
* Allowing leakage of stray light
* Faulty safe-light

Thin patient tissues.

image

Fig. 18.1 Example of a periapical that is too dark with poor contrast.

image

Fig. 18.2 Example of a fogged (blackened) film. Bitewing fogged in the darkroom by inadvertently exposing the upper part of the film to light. The lower part was protected by the operator’s fingers.

Film too pale (Fig. 18.3)

Possible causes

Underexposure owing to:

Faulty X-ray equipment, e.g. timer
Incorrect exposure time setting by the operator
Failure to keep timer switch depressed throughout the exposure

Underdevelopment owing to:

Inadequate time in the developer solution
Developer solution too cold
Developer solution too dilute
Developer solution exhausted
Developer contaminated by fixer

Excessive thickness of patient’s tissues

Film packet back to front (film also marked).

image

Fig. 18.3 Example of a periapical that is too pale with poor contrast.

  Page 210 

Film with inadequate or low contrast (Figs 18.1, 18.2, 18.3)

Possible causes

Processing error owing to:

Underdevelopment (film also pale)
Overdevelopment (film also dark)
Developer contaminated by fixer
Inadequate fixation time
Fixer solution exhausted

Fogging owing to:

Poor storage conditions:
* Allowing exposure to stray radiation
* Too warm
Poor stock control and film used after expiry date
Faulty cassettes allowing the ingress of stray light
Faulty darkroom/processing unit.

Image unsharp and blurred (Fig. 18.4)

Possible causes

Movement of the patient during the exposure (see also Chs 10, 11 and 17)

Excessive bending of the film packet during the exposure (see also Ch. 10)

Poor film/screen contact within a cassette

Film type — image definition is poorer with indirect-action film than with direct-action film

Speed of intensifying screens — fast screens result in loss of detail

Overexposure — causing burn-out of the edges of a thin object

Poor positioning in panoramic radiography (see Ch. 17).

image

Fig. 18.4 Examples of unsharp and blurred films. A As a result of patient movement. B As a result of excessive bending of the film packet during the exposure.

Film marked (Fig. 18.5)

Possible causes

Film packet bent by the operator

Careless handling of the film in the darkroom resulting in marks caused by:

Finger prints
Finger nails
Bending
Static discharge

Processing errors owing to:

Chemical spots
Under fixation — residual silver halide emulsion remaining
Roller marks
Protective black paper becoming stuck to the film
Insufficient chemicals to immerse films fully

Patient biting too hard on the film packet

Dirty intensifying screens in cassettes.

image

Fig. 18.5 Examples of marked films.

A Finger print impression in the emulsion (arrowed).

B Finger nail marks (arrowed).

C Sharply bent film (arrowed) damaging the emulsion.

D Discharge of static electricity (arrowed).

E Fixer splashes on the emulsion before the film was placed in the developer.

F Marks (arrowed) caused by residual emulsion remaining following inadequate fixation (these are usually brown).

  Page 211 
  Page 212 
  Page 213 
  Page 214 

PATIENT PREPARATION AND POSITIONING (RADIOGRAPHIC TECHNIQUE) ERRORS (FIG. 18.6)

These errors can happen whatever image receptor is being used and were described in detail and illustrated in Chapters 10, 11 and 17. They are summarized below and can be divided into intraoral and panoramic technique errors.

image

Fig. 18.6 Two examples of positioning (radiographic technique) errors. A Intraoral – coning off or cone cutting – X-ray tubehead incorrectly positioned, anterior part of image receptor not exposed. B Panoramic — patient too far away from the image receptor — incisor teeth magnified (anteroposterior error) and patient rotated to the LEFT, left molars narrowed, right molars widened (horizontal error).

Intraoral technique errors

These can include:

Failure to position the image receptor correctly to capture the area of interest

Failure to position the image receptor correctly causing it to bend (if flexible) creating geometrical distortion

Failure to orientate the image receptor correctly and using it back-to front

Failure to align the X-ray tubehead correctly in the horizontal plane causing:

coning off or cone cutting
overlapping and geometrical distortion
superimposition

Failure to align the X-ray tubehead correctly in the vertical plane causing:

coning off or cone cutting
oreshortening and geometrical distortion
elongation and geometrical distortion

Failure to instruct the patient to remain still during the exposure with subsequent movement resulting in blurring

Failure to set correct exposure settings (image too dark or too pale — see earlier)

Careless inadvertent use of the image receptor twice.

Panoramic technique errors

These can include:

Failure to remove jewellery

Failure to remove dentures

Failure to remove orthondontic appliances

Failure to remove spectacles

Inappropriate use of a protective lead apron

Failure to ensure the spine is straight

Failure to ensure the incisors are biting on the bite-peg (anteroposterior error)

Failure to use the light beam markers to ensure mid-sagittal plane is vertical and Frankfort plane is horizontal (horizontal and vertical errors)

Failure to instruct the patient to press the tongue against the roof of the mouth

Failure to instruct the patient to remain still throughout the exposure cycle

Failure to set machine height adjustment correctly

Failure to set correct exposure settings (image too dark or too pale — see earlier)

Failure to use the cassette/image receptor correctly.

  Page 215 

QUALITY ASSURANCE IN DENTAL RADIOLOGY

The World Health Organization has defined radiographic quality assurance (QA) programmes as ‘… an organised effort by the staff operating a facility to ensure that the diagnostic images produced by the facility are of sufficiently high quality so that they consistently provide adequate diagnostic information at the lowest possible cost and with the least possible exposure of the patient to radiation’.

Quality control measures are therefore as essential in a general dental practice facility, as they are in a specialized radiography department. This importance of quality is acknowledged in the UK in the Ionising Radiations Regulations 1999 (see Ch. 8) which make quality assurance in dental radiography a mandatory requirement. A section in the 2001 Guidance Notes for Dental Practitioners on the Safe Use of X-ray Equipment is devoted to quality assurance and should be regarded as essential reading. This chapter is based broadly on the recommendations in the 2001 Guidance Notes.

Terminology

The main terms in quality procedures include:

Quality control — the specific measures for ensuring and verifying the quality of the radiographs produced.

Quality assurance — the arrangements to ensure that the quality control procedures are effective and that they lead to relevant change and improvement.

Quality audit — the process of external reassurance and assessment that quality control and quality assurance mechanisms are satisfactory and that they work effectively.

Quality assurance programme

A basic principle of quality assurance is that, within the overall QA programme, all necessary procedures should be laid down in writing and in particular:

Implementation should be the responsibility of a named person

Frequency of operations should be defined

The content of the essential supporting records should be defined and the frequency for the formal checking of such records.

As stated in the 2001 Guidance Notes and implied by the WHO definition, a well-designed QA programme should be comprehensive but inexpensive to operate and maintain. The standards should be well researched but once laid down would be expected to require only infrequent verification or modification. The procedures should amount to little more than ‘written down common sense’. The aims of these programmes, whether using film-based or digital radiography, can be summarized as follows:

To produce diagnostic radiographs of consistently high standard

To reduce the number of repeat radiographs

To determine all sources of error to allow their correction

To increase efficiency

To reduce costs

To ensure that radiation doses to patients and staff are kept as low as reasonably practicable (ALARP).

Quality control procedures for film-based radiography

The essential quality control procedures relate to:

Image quality and film reject analysis

Patient dose and X-ray equipment

Darkroom, image receptors and processing

Working procedures

Staff training and updating

Audits.

Image quality and film reject analysis

Image quality assessment is an important test of the entire QA programme. Hence the need for clinicians to be aware of all the various factors, outlined earlier, that affect image quality and to monitor it on a regular basis. This assessment should include:

  Page 216 

A day-to-day comparison of the quality of every radiograph to a high standard reference film positioned permanently on the viewing screen and an investigation of any significant deterioration in quality.

A formal analysis of film quality, either retrospective or prospective, approximately every six months. The 2001 Guidance Notes recommended the simple three-point subjective rating scale shown in Table 18.2, and shown previously in Chapters 10, 11 and 17, be used for film-based intra-oral and extra-oral radiography.

Based on these quality ratings, performance targets can be set. Suitable targets recommended in the Guidance Notes are shown in Table 18.3 with the advice that practices should aim to achieve these targets within three years of implementing the QA programme. The ‘interim targets’ should be regarded as the minimum achievable standard in the shorter term.

Analysis of all unacceptable films given a rating of 3 sometimes referred to as film reject analysis (see below).

Table 18.2 Subjective quality rating criteria for film-captured images published in 2001 Guidance Notes for Dental Practitioners on the Safe Use of X-ray Equipment

Rating Quality Basis
1 Excellent No errors of patient preparation, exposure, positioning, processing or film handling
2 Diagnostically acceptable Some errors of patient preparation, exposure, positioning, processing or film handling, but which do not detract from the diagnostic utility of the radiograph
3 Unacceptable Errors of patient preparation, exposure, positioning, processing or film handling, which render the radiograph diagnostically unacceptable

Table 18.3 Minimum and interim targets for radiographic quality from the 2001 Guidance Notes

  Percentage of radiographs taken
Rating Target Interim target
1 Not less than 70% Not less than 50%
2 Not greater than 20% Not less than 40%
3 Not greater than 10% Not greater than 10%

Film reject analysis

This is a simple method of identifying all film faults and sources of error and amounts to a register of reject radiographs. To do this, it is necessary to collect all rejected (grade 3) radiographs and record:

Date

Nature of the film fault/error, as shown earlier, e.g.:

a. Film too dark
b. Film too pale
c. Low or poor contrast
d. Unsharp image
e. Poor positioning

Known or suspected cause of the error or fault and corrective action taken (see Table 18.1)

Number of repeat radiographs (if taken)

Total number of radiographs taken during the same time period. This allows the percentage of faulty films to be calculated.

Regular review of film reject analysis records is an invaluable aid for identifying a range of problems, including a need for equipment maintenance, additional staff training as well as processing faults that could otherwise cause unnecessary radiation exposure of patients and staff.

Patient dose and X-ray equipment

One of the aims of QA stated earlier is to ensure that radiation doses are kept as low as reasonably practicable. It is therefore necessary to measure patient doses on a regular basis and compare them against national diagnostic reference levels. To achieve this, X-ray equipment must comply with current recommendations, as described in Chapter 8. These include:

  Page 217 

The initial critical examination and report — carried out by the installer

The acceptance test — carried out by the radiation protection adviser before equipment is brought into clinical use and includes measurement of patient dose

A re-examination report following any relocation, repair or modification of equipment that may have radiation protection implications

Day-to-day checks of important features that could affect radiation protection including:

correct functioning of warning lights and audible alarms
correct operation of safety devices
satisfactory performance of the counterbalance for maintaining the correct position of the tubehead

Written records and an equipment log should be maintained and include:

all installer’s formal written reports describing the checks made, the results obtained and action taken
results of all equipment checks in chronological order
details of all routine or special maintenance

The Ionizing Radiation (Medical Exposure) Regulations 2000 require that an up-to-date inventory of each item of X-ray equipment is maintained, and available, at each practice and contains:

name of the manufacturer
model number
serial number or other unique identifier
year of manufacture
year of installation

Compliance with the recommendations in Ch. 9 of Recommended standards for the routine performance testing of diagnostic X-ray imaging systems, Report 91 of the Institute of Physics and Engineering in Medicine (IPEM) published in 2005.

Darkroom, image receptors and processing

Darkroom

The QA programme should include instructions on all the regular checks that should be made, and how frequently, with all results recorded in a log. Important areas include:

General cleanliness (daily), but particularly of work surfaces and film hangers (if used).

Light-tightness (yearly), by standing in the darkroom in total darkness with the door closed and safelights switched off and visually inspecting for light leakage

Safelights (yearly), to ensure that these do not cause fogging of films. Checks are required on:

Type of filter — this should be compatible with the colour sensitivity of film used, i.e. blue, green or ultraviolet (see Ch. 6)
Condition of filters — scratched filters should be replaced
Wattage of the bulb — ideally it should be no more than 25W
Their distance from the work surface — ideally they should be at least 1.2m (4ft) away
Overall safety (i.e. their fogging effect on film) — the simple quality control measure for doing this is known as the coin test:

1. Expose a piece of screen film in a cassette to a very small even exposure of X-rays (so-called flash exposure) to make the emulsion ultra-sensitive to subsequent light exposure
2. In the darkroom, remove the film from the cassette and place on the worksurface underneath the turned-off safelight
3. Place a series of coins (e.g. seven) in a row on the film and cover them all with a piece of card
4. Turn on the safelight and then slide the card to reveal the first coin and leave for approximately 30 seconds
5. Slide the card along to reveal the second coin and leave again for approximately 30 seconds
6. Repeat until all the coins have been revealed
7. Process the film in the normal way.

A simulated result is shown in Figure 18.7. Fogging (blackening) of the film owing to the safelight will then be obvious when compared to the clear area protected by the coin. The part of the film adjacent to the first coin will have been exposed to the safelight for the longest time and will be the darkest. In practice, the normal film-handling time under the safelight can be measured and the effect of safelight fogging established.

image

Fig. 18.7 A simulated coin test result. The film, with seven coins on it, has been gradually uncovered every 30 seconds. The coin-covered part of the film remains white while the surrounding film is blackened or fogged. The longer the film is exposed to the safelight the darker it becomes.

(Kindly provided by Mr N. Drage.)

Note. The coin test can also be used to assess the amount of light transmission through the safety glass of automatic processors by performing the test within the processor under the safety glass under normal daylight loading conditions.

Image receptors

The QA programme requires written information, usually obtained from the suppliers, on film speed, expiry date and storage conditions as well as details regarding the maintenance and cleaning instructions of cassettes and/or digital image receptors. Typical requirements could include:

X-ray film

This requires:

Ideal storage conditions — cool, dry and away from all sources of ionizing radiation — as recommended by the manufacturers

Strict stock control with records to ensure usage before the expiry date

Careful handling.

Cassettes

These require:

Regular cleaning of intensifying screens with a proprietary cleaner

Regular checks for light-tightness, as follows:

1. Load a cassette with an unexposed film and place the cassette on a window sill in the daylight for a few minutes
2. Process the film — any ingress of light will have fogged (darkened) the film (see Fig. 18.8(i))

Regular checks for film/screen contact, as follows:

1. Load a cassette with an unexposed film and a similar sized piece of graph paper
2. Expose the cassette to X-rays using a very short exposure time
3. Process the film — any areas of poor film/screen contact will be demonstrated by loss of definition of the image of the graph paper (see Fig. 18.8(ii))

A simple method of identification of films taken in similar looking cassettes, e.g. a Letraset letter on one screen.

image

Fig. 18.8(i) Radiograph from a faulty cassette being checked for light-tightness. The light that has got into the cassette has blackened one side of the film.

image

Fig. 18.8(ii) Examples of the radiographs following the graph paper test for film/screen contact. A Good film/screen contact — the fine detail of the graph paper is evident over the whole film. B Poor film/screen contact — note the loss of detail in several areas.

Processing

The QA programme should contain written instructions about each of the following:

Chemical solutions

These should be:

Always made up to the manufacturers’ instructions taking special precautions to avoid even trace amounts of contamination of the developer by the fixer, e.g. always fill the fixer tank first so that any splashes into the developer tank can be washed away before pouring in the developer

Always at the correct temperature

Changed or replenished regularly — ideally every 2 weeks — and records should be kept to control and validate these changes

Monitored for deterioration. This can be done easily using radiographs of a step-wedge phantom:

1. Make a simple step-wedge phantom using the lead foil from inside intra-oral film packets, as shown in Figure 18.9(i)
2. Radiograph the step-wedge using known exposure factors
3. Process the film in fresh solutions to produce a standard reference film
4. Repeat, using the same exposure factors, every day as the solutions become exhausted
5. Compare each day’s film with the standard reference film to determine objectively any decrease in blackening of the processed film which would indicate deterioration of the developer (see Fig. 18.9(ii))
6. Record the results.
image

Fig. 18.9(i) A simple step-wedge phantom constructed using pieces of lead foil taped to a tongue spatula.

image

Fig. 18.9(ii) A The standard reference film of the step-wedge phantom on DAY ONE processed using newly made-up chemical solutions. B Test film processed in chemical solutions 1 week old — note the reduced amount of blackening of the second film owing to the weakened action of the developer.

Processing equipment

Manual processing requires the use of accurate timers, thermometers and immersion heaters. Instructions on their proper use should be provided.

Automatic processors require regular replenishment of chemical solutions and regular cleaning, especially of the rollers. All cleaning procedures should be written down including how often they should be carried out.

Record log confirming that all cleaning procedures have been carried out should be kept.

Working procedures

These include:

Local rules — required in the UK under the Ionising Radiations Regulations 1999 (see Ch. 8). These rules should contain the procedural and operational elements that are essential to the safe use of X-ray equipment, including guidance on exposure times, and as such should contain much of what is relevant to the maintenance of good standards in QA.

Employers’ written procedures — required in the UK under the Ionising Radiation (Medical Exposure) Regulations 2000 (see Ch. 8).

Operational procedures or systems of work — these include written procedures that provide for all actions that indirectly affect radiation safety and diagnostic quality, e.g. instructions for the correct preparation and subsequent use of processing chemicals (as explained earlier).

Procedures log — the QA programme should include the maintenance of a procedures log to record the existence of appropriate Local Rules and Employers’ Written Procedures, together with a record of each occasion on which they are reviewed or modified (ideally every 12 months).

Staff training and updating

As mentioned in Chapter 8, it is a legal requirement under the Ionising Radiation (Medical Exposure) Regulations 2000 that all practitioners and operators are adequately trained and that continuing professional development (CPD) is undertaken. The details of the training required in the UK are given in Chapter 8. The QA programme should incorporate a register of all staff involved with any aspect of radiography and should include the following information:

Name

Responsibility

Date, nature and details of training received

Recommended date for a review of training needs.

Audits

Each procedure within the QA programme will include a requirement for written records to be made by the responsible person at varying intervals. In addition, the person with overall responsibility for the QA programme should check the full programme at intervals not exceeding 12 months. This is an essential feature of demonstrating effective implementation of the programme. Clinical audits may include:

The QA programme and associated records

The justification and authorisation of radiographs

The appropriateness of requests/investigations

The clinical evaluation of radiographs.

QUALITY CONTROL PROCEDURES FOR DIGITAL RADIOGRAPHY

The overall quality control procedures for digital radiography are similar, and in some instances identical, to those required for film-based radiography. They relate to:

Image quality assessment

Patient dose and X-ray equipment

Image processing and manipulation

Working procedures

Staff training and updating

Audits.

Quality control procedures for digital image quality assessment and image processing and manipulation are currently not well defined or documented, although some of the recommendations in the 2005 IPEM Report 91, mentioned earlier, apply to dental digital imaging. Theoretically image quality overall should be improved as:

  Page 221 

Chemical processing errors are eliminated

Most exposure errors can be compensated for by using computer software particularly when using phosphor plates with their wide latitude, although overexposure of solid-state sensors can cause blooming (see Ch. 7 and Fig. 7.14).

However, radiographic technique errors are just as important as in film-based radiography and need to be monitored. Clinicians using digital radiography are still required to have a quality assurance (QA) programme. The 2001 Guidance Notes make the additional recommendation that because of the ease with which digital radiographs can be retaken, it is essential to ensure that all retakes are properly justified, recorded and included in QA statistics.

Footnote

The requirement for quality assurance and quality control measures in general dental practice applies equally to specialized radiography departments. However, in view of the cost implications, the expensive, sophisticated equipment available for precise quality assurance measurements and accurate monitoring in X-ray departments are often inappropriate to general practice. The practical suggestions in this chapter for film-based radiography, based on the 2001 Guidance Notes, are designed to satisfy the WHO definition by bringing an element of objectivity to quality assurance in practice, but at the same time being simple, easily done and inexpensive.

  Page 222